Thermoluminescence dosimetric characteristics of fabricated germanium (Ge) doped optical fibres for electron beams dosimetry: a preliminary study

Zabariah Zakaria, and Mohd Zahri Abdul Aziz, and Noramaliza Mohd Noor, (2020) Thermoluminescence dosimetric characteristics of fabricated germanium (Ge) doped optical fibres for electron beams dosimetry: a preliminary study. Sains Malaysiana, 49 (8). pp. 1981-1985. ISSN 0126-6039

[img]
Preview
PDF
303kB

Official URL: http://www.ukm.my/jsm/malay_journals/jilid49bil8_2...

Abstract

The basic dosimetric favourable responses of tailored fabricated germanium (Ge) doped cylindrical optical fibres, make use of 2.3% mol have been extensively studied with clinical electron beams irradiation in terms of dose linearity, reproducibility, fading signals, minimum detectable dose (MDD), energy- and field size dependence. The irradiation was performed at the Radiotherapy Unit, Advanced Medical and Dental Institute (AMDI), which utilizes Elekta Synergy® linear accelerator (LINAC) at 6-, 9-, 12- and 15 MeV electron beam energies at specific doses ranging from 1 to 5 Gy. Thermoluminescence (TL) signals exhibited a linear dose-response overdose ranges, mean reproducibility with a coefficient of variation (CV) of better than 10% and no dependency with different field sizes at p > 0.05. The MDD values were typically 3.51 to 4.13 mGy. The minimum TL fading of the fabricated Ge-doped cylindrical optical fibres was reported favourably for 9 MeV electron beam at day of 74.

Item Type:Article
Keywords:Dose linearity; Optical fibres; Thermoluminescence (TL)
Journal:Sains Malaysiana
ID Code:15791
Deposited By: ms aida -
Deposited On:17 Nov 2020 04:42
Last Modified:22 Nov 2020 15:30

Repository Staff Only: item control page