Ercan, Zeynep Eylül (2021) Correlation of corneal thickness measurements between tono-pachymeter and scheimpflug-placido topography. Medicine & Health, 16 (2). pp. 161-167. ISSN 2289-5728
|
PDF
337kB |
Official URL: https://medicineandhealthukm.com/toc/16/2
Abstract
Central corneal thickness (CCT) measurements are important for diagnosis,treatment, and surgery planning in ophthalmology. The purpose of this study was to see whether CCT measurements taken with Tono-pachymeter and Scheimpflug-Placido Topography had any significant differences. Tono-pachymeter and topography CCT measurements were taken (n=400). Inter-measurement agreement between them was determined using Bland-Altman Plot analysis. Age groups were also formed as group 1 (aged 18-50 years, 94 males, 106 females) and group 2 (age >51 years, 100 males, 100 females). Mean CCTs measured by Tono-pachymeter and topography were 563.77 +±26.43 and 560.88 + 26.341 microns. Bland-Altman Plot analysis showed in total, 13 were above the upper limit and 5 were under the minimum limit of agreement with regression analysis showing no significant relationships (p=0.213). Group 1 had 7 above and 2 below from the limits of agreement. Group two had 9 above and 2 below from the limits of agreement. Both groups showed insignificant differences between devices (p=0.07 and p=0.86). Tono-pachymeter and Scheimpflug-Placido Topography give reliable CCT results within each other. However, since the limit of agreement ranges can still affect one-to-one patient evaluations, we recommend clinics that use these devices to not interchange measurements in practice.
Item Type: | Article |
---|---|
Keywords: | Corneal pachymetry; Corneal topography; Cornea; Ophthalmological diagnostic techniques |
Journal: | Medicine & Health |
ID Code: | 18274 |
Deposited By: | ms aida - |
Deposited On: | 24 Mar 2022 00:16 |
Last Modified: | 25 Mar 2022 01:48 |
Repository Staff Only: item control page