Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector

Samat, S.B. and Priharti, W. (2015) Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector. Sains Malaysiana, 44 (5). pp. 761-764. ISSN 0126-6039

[img]
Preview
PDF
264kB

Official URL: http://www.ukm.my/jsm/malay_journals/jilid44bil5_2...

Abstract

A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector’s energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector’s Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10−1 keV, respectively. The Fano factor value agrees within 95% confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer’s specification.

Item Type:Article
Keywords:Fano factor; FWHM; gamma-ray spectrometry; pre-amplifier noise
Journal:Sains Malaysiana
ID Code:8953
Deposited By: ms aida -
Deposited On:26 Jul 2015 14:59
Last Modified:14 Dec 2016 06:48

Repository Staff Only: item control page